Rapid focus map surveying for whole slide imaging with continues sample motion

July 06, 2017 Β· Declared Dead Β· πŸ› Optics Letters

πŸ‘» CAUSE OF DEATH: Ghosted
No code link whatsoever

"No code URL or promise found in abstract"

Evidence collected by the PWNC Scanner

Authors Jun Liao, Yutong Jiang, Zichao Bian, Bahareh Mahrou, Aparna Nambiar, Alexander W. Magsam, Kaikai Guo, Yong Ku Cho, Guoan Zheng arXiv ID 1707.03039 Category cs.CV: Computer Vision Citations 45 Venue Optics Letters Last Checked 6 months ago
Abstract
Whole slide imaging (WSI) has recently been cleared for primary diagnosis in the US. A critical challenge of WSI is to perform accurate focusing in high speed. Traditional systems create a focus map prior to scanning. For each focus point on the map, sample needs to be static in the x-y plane and axial scanning is needed to maximize the contrast. Here we report a novel focus map surveying method for WSI. The reported method requires no axial scanning, no additional camera and lens, works for stained and transparent samples, and allows continuous sample motion in the surveying process. It can be used for both brightfield and fluorescence WSI. By using a 20X, 0.75 NA objective lens, we demonstrate a mean focusing error of ~0.08 microns in the static mode and ~0.17 microns in the continuous motion mode. The reported method may provide a turnkey solution for most existing WSI systems for its simplicity, robustness, accuracy, and high-speed. It may also standardize the imaging performance of WSI systems for digital pathology and find other applications in high-content microscopy such as DNA sequencing and time-lapse live-cell imaging.
Community shame:
Not yet rated
Community Contributions

Found the code? Know the venue? Think something is wrong? Let us know!

πŸ“œ Similar Papers

In the same crypt β€” Computer Vision

πŸŒ… πŸŒ… Old Age

Fast R-CNN

Ross Girshick

cs.CV πŸ› ICCV πŸ“š 27.7K cites 11 years ago

Died the same way β€” πŸ‘» Ghosted